# # Dual Grating MOS Lamps # # Template for taking MOS flat wavelength calibration lamps in # dual mode. # # Just 1 lamp per mask is needed # # R. Pogge, OSU Astronomy Dept. # pogge@astronomy.ohio-state.edu # 2011 Feb 19 [last mod] # #--------------------------------------------------------------------------- calmode Instrument: instconfig dual grating slitmask ID543210 Exec: red nimgs 1 blue nimgs 1 NeHgLamp: # Ne + Hg Lamps - wait 30s for Hg to warm up # ND1.5 filter in both channels. 3 images per side print Ne+Hg Lamp Spectra lamp hg ne on print Waiting 30s for the Hg lamp to warm up sleep 30 comp Ne + Hg[Ar] Lamps blue exptime 2 blue filter ND1.5 red exptime 1 red filter ND1.5 go lamp off KrXeLamp: # Kr+Xe Lamps - Red: ND1.5 3x1s, Blue: Clear 3x15s print Kr+Xe Lamp Spectra lamp xe kr on comp Xe+Kr Lamps blue exptime 15 blue filter clear red exptime 1 red filter nd1.5 go lamp off ArLamp: # Argon Lamp - Red: ND1.5 1s, Blue: Clear 30s print Ar Lamp Spectra lamp ar on comp Ar Lamp blue exptime 30 blue filter clear red exptime 1 red filter nd1.5 go lamp off end